Deutsch English
 home Products | Production | Development | Test | Contact | About us | E-Mail 
 

Anaesthesia Monitor BIO-3A

Pulse oximeter with ECG unit BIOSCOPE-D.

 BIO-3A
 
 BIOSCOPE-C
 
 BIOTREND
 
 RECOMED
 
 URO-2050
 
 No longer
 deliverable:
 
 BIO-3
 
 VITASCOPE-A
This device stands out by the combination of a pulse oximeter and a separate usable (removable from the monitor) BIOSCOPE-D. Pulse oximeter and BIOSCOPE-D are united to a small compact monitor. The monitor could be used as a desktop model or fastened at an infusion stand by a tripod mount.
 
The anaesthesia monitor BIO-3A is suitable both for mains and battery operation. It also has built-in rechargeable batteries besides an external power supply. The operation time with fully loaded batteries is about eight hours. The batteries will be loaded with the monitors built-in charger. On the one hand the BIO-3A could be used in an uninterrupted mains operation mode at a permanent location, on the other hand a mobile use is also possible.
 
The monitor BIO-3 is not developed for bed-sided use. It is suitable for operations while medical specialist staff is in the proximity and can directly react to a triggered alarm.
 
Anaesthesia Monitor BIO-3A with BIOSCOPE-D.
Click on the picture
to enlarge it.
Application range:
  • alarm function for SpO2 and heart rate
  • acoustic and optical signal for heart rate
  • due to built-in rechargeable batteries independent of the mains, therefore also suitable for transportable use
  • no small single devices with unsafe location
  • nevertheless ECG monitor BIOSCOPE-D also separate usable
  • data storage on PC
Standard accessories:
  • ECG patient cable
  • oximeter finger sensor for adults
  • power supply
Optional accessories:
  • tripod mount
  • PC software for documentation and evaluation: signal view of ECG and plethysmogramm, trend HR and SpO2, save, print, data export

Technical data:

 
ECG by BIOSCOPE-D
one-channel recording
heart-rate
   - measurement range
30 ... 240 bpm
heart-rate
   - measurement accuracy
30 ... 150 bpm: ±1 bpm
151 ... 240 bpm: ±2 bpm
sensitivity 0.25 ... 8 mV
(automatic level control in 4 steps)
bandwidth 0.5 ... 40 Hz (3dB)
time constant 0.33 sec
input impedance 2 MOhm
suppression of mains hum
marking of pacemaker pulses
average heart rate displayed by the LED display
in case ECG is not connected the
averaged pulse rate will be displayed (derived by finger sensor)
alarm limits adjustable over the whole measurement range
 
SpO2 by Pulse-Oximeter
oxygen saturation
   - measurement range
45 ... 99 %
oxygen saturation
   - accuracy
70 ... 99 %: < 2 %
< 70 %: not specified
pulse rate
   - measurement range
30 ... 240 bpm
alarm limits adjustable over the whole measurement range
(lower limit value only)
bar graph display for assessment of quality of pulse signal
sensor universal hinged finger sensor for adults
(other sensors on request)
 
General
power supply battery and mains operation
mains voltage 230V ±10%, 50 Hz
running time in battery mode 6 ... 8 h permanent operation
electrical safety The monitor BIO-3 agrees with:
EN 60601-1 (06.95), EN 60601-1-2 (11.2001),
ECG - BF, SpO2 - BF, defi protected,
class IIb acc. MDD 93/42/EEC
about medical products (06/93), CE 0123
dimension approx. 160 x 180 x 120 mm (w x d x h)
weight approx. 1.3 kg (incl. BIOSCOPE-D, without accessories)
Currently not available !

The service for BIO-3A is continued.
 
Modifications reserved.

Optional PC software for BIO-3A
PC software BIO-3A.
Click on the picture
to enlarge it.
The optional available PC software BIO-3A is the perfect tool to document the data derived from the anaesthesia monitor BIO-3A. So you can see the signals of ECG and plethysmogramm on a larger screen of a PC and watch the trend of heart rate and SpO2 in real time. Certainly it is possible to store the transmitted data for a later processing. The software has a built-in printing function for your documentation. Also is a data export possible.
Minimum requirements for the used PC:
Operating sytem: Windows 2000 or Windows XP
Memory: 64 MByte RAM
Available space on hard disk: 10 MByte plus size for storing measured values
Drive: CD drive
Screen resolution: at least 1024 x 768 pixel
Interface: serial interface (RS232)

Please send an email to sales@recomedical.com to ask for pricing.
 
Note:
Operating instructions and software updates you can find on our download page.
 
Imprint